Power Frequency Magnetic Field-Testing Labs in Uusimaa

A list of EMC testing labs that provide Power Frequency Magnetic Field Testing in Uusimaa, Finland. Power frequency magnetic field testing evaluates the immunity of electronic equipment to low-frequency magnetic fields generated by AC power systems and high-current electrical infrastructure. These magnetic fields typically originate from transformers, busbars, switchgear, and power cables operating at mains frequencies. We have listed leading power-frequency magnetic field-testing labs worldwide and made them searchable by location and capabilities.
Selected Filters
  • Located in: Finland
  • Serves: Finland
  • State: Uusimaa
  • Test Services: Power Frequency Magnetic Field Testing
  • Reset All
  • Would you like an EMC Testing Quote from Uusimaa?

    Tell us your EMC Testing requirement and let us get you quotes from multiple companies.

    Get Quotes Now
Advertisement

Power Frequency Magnetic Field Testing

Power frequency magnetic field testing evaluates how electronic equipment responds to magnetic fields generated by power systems operating at 50 Hz or 60 Hz. High currents flowing through power distribution conductors, transformers, and switchgear produce low-frequency magnetic fields that can couple into nearby electronic circuits through inductive coupling, inducing unwanted voltages or currents in sensitive signal paths.

The test is performed by placing the device under test (DUT) inside a calibrated magnetic field generated by single-axis induction coils or Helmholtz coil systems. These coils are driven by a controlled power frequency current source, producing a uniform sinusoidal magnetic field within the test volume. The magnetic field strength is monitored and calibrated using magnetic field probes to ensure the required field level is achieved at the DUT location.

Testing is typically conducted by applying the magnetic field sequentially along three orthogonal axes (X, Y, and Z) to evaluate susceptibility to magnetic coupling from different orientations. The DUT operates in normal functional modes while exposure levels are increased to specified field strengths, allowing engineers to observe any functional degradation, performance deviation, or loss of operation.

Power frequency magnetic field immunity testing is defined in IEC 61000-4-8, which specifies the coil configuration, calibration procedures, field uniformity requirements, and test severity levels. Standard test levels commonly range from 1 A/m to 100 A/m, with higher levels used for industrial or substation environments. These tests verify the immunity of equipment installed in environments with strong power frequency magnetic fields, such as industrial facilities, power plants, and electrical substations.

EMC Directory is the largest online directory of EMC Testing labs. We have listed leading power-frequency magnetic field testing labs worldwide and made them searchable by their location and capabilities.

Get EMC TestingQuotes from Uusimaa

Tell us your EMC Testing requirement and let us get you quotes from multiple companies.

Get Quotes Now

Get EMC Testing Quotes from Uusimaa

Tell us your EMC Testing requirement and let us get you quotes from multiple companies.

Get Quotes Now
directory ../ test_services=Power Frequency Magnetic Field Testing; 0 directory emc-testing-services
True 2026/05_21